Recall Z-0522-2012 ยท LLZ System, Image Processing, Radiological

Siemens Medical Solutions USA, Inc

syngo.plaza Image processing system

DateJan 11, 2012 (initiated Sep 21, 2011)
StatusTerminated
ReasonPotential malfunction when using syngo.plaza. On syngo.plaza workplaces large studies can be loaded into the viewer if transferred from a DICOM node or imported even while image transfer is not yet completed. In these situations more images of that study may exist which are not shown in the syngo.plaza viewer.
Root causeSoftware design
ActionSiemens Medical Solutions USA sent a Customer Safety Advisory Notice dated March 10, 2011 to the affected customer. The letter identified the product the problem and the action needed to be taken by the customer. The letter informed the customer of the potential issues and provides additional information on the implemented resolution. For further questions please call (610) 219-4834.
Quantity1
DistributionNationwide Distribution New York
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K093612SYNGO.PLAZA
Siemens AG Medical Solutions, Mar 4, 2010