Recall Z-0514-2012 ยท LLZ System, Image Processing, Radiological

Siemens Medical Solutions USA, Inc

syngo Imaging XS, all versions Radiological image processing system

DateJan 11, 2012 (initiated Sep 20, 2011)
StatusTerminated
ReasonUnintended behavior when using syngo Imaging XS, all versions. The "Rename" functionality can cause an unintended patient merge if the renamed patient is sent to another system.
Root causeLabeling False and Misleading
ActionSiemens Medical Solutions USA, Inc sent a Customer Safety Advisory Notice letter dated August 3, 2011 to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. If you detect patient data inconsistencies in your system caused by this problem, please contact Siemens Service for support. For further questions please call (610) 219-4834.
Quantity38
DistributionNationwide Distribution including the states of CA, FL, GA, IN, IA, KS, MN, MO, NE, NV, NJ, NY, NC, OH, PA, TN, TX, and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082430SYNGO IMAGING XS
Siemens AG Medical Solutions, Nov 14, 2008