Recall Z-0515-2012 ยท LLZ System, Image Processing, Radiological
Siemens Medical Solutions USA, Inc
syngo Imaging XS, versions VA70A and VA70B Client/Server Installations Radiological image processing system
| Date | Jan 11, 2012 (initiated Sep 19, 2011) |
| Status | Terminated |
| Reason | Unintended behavior when using syngo Imaging XS, versions VA70A and VA70B Client/Serve installations. Loading additional images/series from Patient Browser using the thumbnails can cause an unintended deletion of images/series. |
| Root cause | Software design |
| Action | Siemens Medical Solutions USA, Inc sent a Customer Safety Advisory letter dated August 24, 2011, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. If you detect patient data inconsistenceis in your system caused by this problem, please contact Siemens Service for support. For further questions please call (610) 219-4834. |
| Quantity | 12 |
| Distribution | Nationwide Distribution including the states of CA, IN, IA, NV, NE, NY, OH, TN and TX. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K082430 | SYNGO IMAGING XS Siemens AG Medical Solutions, Nov 14, 2008 |