Recall Z-0523-2012 ยท MUE Full Field Digital, System, X-Ray, Mammographic

Siemens Medical Solutions USA, Inc

MAMMOMAT Inspiration Full field digital mammographic x-ray system

DateJan 11, 2012 (initiated Sep 15, 2011)
StatusTerminated
ReasonLowering the swivel arm with "object table" while an operator's or patient's body parts are underneath the "object table" may cause the "object table" to collide with and injure the operator or patient. Therefore attention should be paid when adjusting "object table" height to avoid a possible risk of injury to the operator or the patient. The movement of the swivel arm may also cause an injur
Root causeUse error
ActionSiemens Medical Solutions USA, Inc sent a Safety Advisory Notice dated January 8, 2011, to all affected customers. The letter identified the product, the problem and the action needed to be taken by the customer. The notification included an Addendum to the User Manual. We appreciate your understanding and cooperation with this safety advisory and ask you to immediately instruct your operating personnel accordingly. Please ensure that this safety advisory is placed in the System's Operator Manual. Your operating personnel should maintain awareness over an appropriate defined period. Furt
Quantity12
DistributionNationwide Distribution including the states of FL, MD, MI, NE, NY, NC, ND, OH, and SC.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.