Recall Z-0388-2012 ยท LLZ System, Image Processing, Radiological

Vital Images, Inc.

Vitrea (available in various configuration as Vitrea Enterprise Suite). Vitrea Enterprise Suite 1.2, Vitrea Enterprise Suite 1.3, Vitrea enterprise Suite 6.0 and Vitrea Enterprise Suite 6.1. A medical diagnostic system that allows the processing, review, analysis, communication and media interchange of multi-dimensional digital images acquired from variety of imaging devices.

DateJan 11, 2012 (initiated Nov 11, 2011)
StatusTerminated
ReasonVital Images has found two issues that may display incorrect Standard Update Values (SUV) when viewing certain PET/CT images. This issue affects the VitreaCore PET/CT feature for Vitrea Enterprise Suite 1.2, 1.3, 6.0, 6.1, including all updates to those versions.
Root causeSoftware design
ActionVital Images, Inc. sent an URGENT DEVICE RECALL letter dated November 11, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were advised to inform all potential users of the issues. Customers were to be contacted by their Service Account Manager to discuss and schedule resolutions. For questions customers were to contact Vital Images Customer Support with any questions toll free at 1-800-208-3005, in Europe +31 (0) 70 4135801, in Asia t +86 10 8588 1800, or online at support@vitalimages.com.
Quantity338 (233 US)
DistributionWorldwide Distribution - USA including AL, AK, AZ, AR, CA, CO, CT, DC, FL, GA, IL, IN, IA, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NJ, NM, NY, NC, ND, OH, OK OR, PA, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, and WY and the countries of IRELAND, FRANCE, UNITED KINGDOM, SWITZERLAND, AUSTRALIA, S
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K061624VITREA, VERSION 3.9
Vital Images, Inc., Jun 27, 2006
K071331VITREA VERSION 4.0
Vital Images, Inc., May 25, 2007