TrueBeam and TrueBeam STx, Model Number: H19, Ref/FSCA identifier: CP-06381 are intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.

DateDec 2, 2011 (initiated Oct 31, 2011)
StatusTerminated
ReasonAn anomaly has been identified with the TrueBeam and TrueBeam STx systems where, under certain tuning conditions, the electron beam emerging from the bend magnet may have an elongated spot shape.
Root causeDevice Design
ActionOn 11/04/2011, an Urgent Medical Device Correction Letter was distributed to all affected users, with a description of the problem and user corrective action steps. Varian is developing a corrective action for this issue. A customer service representative will contact you when the correction is available to schedule installation on your system. Questions regarding the letter should be directed to the Varian Oncology Help Desk Contact Information at USA and Canada: 1-888-827- 4265 or Europe +41-41-749-8844.
DistributionWorldwide distribution including the US and Internationally.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111106TRUEBEAM
Varian Medical Systems, Inc., Aug 18, 2011