Recall Z-0585-2012 ยท KPR System, X-Ray, Stationary

GE Healthcare, LLC

GE Healthcare Discovery XR650, Digital Radiographic Systems. Revolution XR/d Digital Radiographic Imaging System with Image Pasting and Autopositioning is indicated for use in generating radiographic images of human anatomy. This device is not intended for mammographic applications.

DateJan 11, 2012 (initiated Nov 22, 2011)
StatusTerminated
ReasonGE Healthcare has become aware of a potential issue associated with the patient barrier rotating arm that is a part of the Definium 8000 and Discovery XR650 image pasting systems that may impact patient and user safety. The patient barrier rotating arms that are used to position and stabilize the patient during image pasting procedures may fail to retain residual friction at the arm pivot. Thi
Root causeDevice Design
ActionThe firm, GE Healthcare, sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated November 17, 2011 to its customers. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. The customers were instructed to ensure that all potential users in their facility are made aware of this safety notification and the recommended actions; discontinue use of the patient barrier and remove from service (away from patients) if either one the rotating arm locking levers is damaged (does not hold the rotating arm in place when engaged);
Quantity229
DistributionWorldwide distribution: USA (nationwide) including states of: AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA, HI, ID,IL, IN, IA, KS, KY, LA,ME, MD, MA, MI, MN, MS, MO, MT, NH, NJ, NM, NY,NC, OH, OR, PA, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, and WY; and countries of: VENEZUELA, UNITED KINGDOM, UNITED
FDA recordOpen on fda.gov

510(k) clearances named in this recall