Recall Z-0463-2012 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

Smith & Nephew, Inc. Endoscopy Division

Smith & Nephew 5.5mm DYONICS BONECUTTER ELECTROBLADE Resector Catalog Number: 7205962.

DateJan 11, 2012 (initiated Nov 7, 2011)
StatusTerminated
ReasonManufacturing issue: The distal tip of the outer sheath (return electrode) may detach and become loose in the joint during use.
Root causeProcess control
ActionSmith & Nephew Inc. Endoscopy Division notified facilities on 11/7/11 via an "Urgent - Product Recall 1st Notification - Urgent" letter and/or phone. The letter identified the affected product and the reason for the recall. It also discussed the potential risks, customer actions, and instructions for the return of the product. Customers were to check their inventory and locate any of the affected product. They were to also complete the requested contact information and return the notification letter. In addition, customers were to contact Smith & Nephew Endoscopy Division Returns Group at
Quantity1743 units
DistributionWorldwide Distribution -- USA, Australia, Austria, Belgium, Canada, Dubai, Finland, France, Germany, Norway, South Africa, Switzerland, United Kingdom, and Venezuela.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K012314DYONICS ELECTROBLADE RESECTOR
Smith & Nephew, Inc., Feb 1, 2002