Recall Z-0333-2012 ยท LWS Implantable Cardioverter Defibrillator (Non-Crt)

Boston Scientific CRM Corp

LATITUDE Patient Management System, Models 6441*, 6442*, 6465*, 6488 (USA). There are two software applications: the programmer software and the LATITUDE remote monitoring system software. The Model 2868 Software Application is loaded onto the Model 3120 programmers. These programmers are used during in-office follow-ups. The software version affected domestically is version 1.05 for COGNIS an

DateDec 1, 2011 (initiated Oct 12, 2011)
StatusTerminated
ReasonCurrent Boston Scientific defibrillator systems measure shock lead impedances up to 200 ohms, and export results to the programmer and LATITUDE monitoring system via the Daily Measurements feature. A measured shock lead impedance that is greater than 125 ohms or less than 20 ohms will generate an alert message on programmer screens and a Red Alert within LATITUDE (if activated). However, the mea
Root causeSoftware design
ActionBoston Scientific sent an "IMPORTANT DIAGNOSTIC INFORMATION" letter dated October 2011 to all affected customers. The letter described the product, problem and actions to be taken by the customers.
Quantity25,300 (14,700 US; 10,600 OUS)
DistributionWorldwide Distribution-USA (nationwide) including DC and Puerto Rico and the countries of AUSTRIA, BELGIUM, DENMARK, FINLAND, FRANCE, GERMANY, IRELAND, ITALY, NETHERLANDS, NORWAY, PORTUGAL, SPAIN, SWEDEN, SWITZERLAND, UNITED KINGDOM, CANADA, AUSTRALIA, NEW ZEALAND, CYPRUS, CZECH REPUBLIC, GREECE, GU