ClearanceBrief
Recall Z-0197-2012 ยท DZJ Driver, Wire, And Bone Drill, Manual

Stryker Instruments Div. of Stryker Corporation

Right/Left IBO Blade Set, for use with Stryker TPS Reciprocating Saw, REF 5100-37-903, Stryker Instruments, Kalamazoo, MI. Product Usage: The intended use of these devices is to cut bone and hard tissue in maxillofacial procedures and have been specifically designed for osteotomies of the mandibular ramus.

DateNov 16, 2011 (initiated Oct 18, 2011)
StatusTerminated
ReasonThere is a potential for these blades to break where the blade is welded on to the arbor. This hazard can result in the following harms: tissue damage due to the broken blade being aspirated or swallowed, additional surgery to remove a broken blade, infection due to an unretrieved device component, complications associated with increased time under anesthesia to retrieve a broken blade, osteomyeli
Root causeDevice Design
ActionThe firm, Stryker, sent an "URGENT MEDICAL DEVICE RECALL NOTIFICATION" letter dated October 18, 2011 to Risk Managers, Office Managers, and Oral Surgeons. The letter described the products under recall, the problem and a series of actions to be taken. Customers were asked to immediately locate recalled products and quarantine. Products were to be shipped back to Stryker. The letter included a Business Reply Form to be faxed back to Stryker Instruments Regulatory Department, 866-521-2762. The letter instructs to forward this notification to all affected locations if the recalled products we
Quantity1,086 units
DistributionWorldwide Distribution - USA (nationwide) and the countries of Brazil, Canada, China, Germany, Greece, Hong Kong, India, Jordan, Miramar, Saudi Arabia, Singapore, Spain, Switzerland, United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K040369STRYKER CONSOLIDATED OPERATING ROOM EQUIPMENT (CORE) SYSTEM
Stryker Instruments, Apr 1, 2004