Recall Z-0280-2012 ยท DHN Antinuclear Antibody, Indirect Immunofluorescent, Antigen, Control

Inova Diagnostics Incorporated

Inova Diagnostic Inc., NOVA Lite ANA KSL Kit Part Number: 708390. in vitro diagnostic.

DateNov 23, 2011 (initiated Sep 26, 2011)
StatusTerminated
ReasonThe NOVA Lite ANA KSL kit label has the wrong fill volume for the IFA System Negative Control (P/N 508186) with 1mL. The component in the kit is correctly labeled and filled with the 0.5mL fill volume.
Root causeProcess change control
ActionOn 9/26/11, Inova Diagnostics forwarded an Urgent Advisory Notice and Product Advisory Notice Fax Back Form to all their customers who purchased the Nova Lite Ana KSL, Product Code: 708390. The letter provides the customers with an explanation of the problem identified and an action to be taken. Customers were instructed to complete the attached fax back form indicating how many of these kits they have in their inventory, quantity destroyed, method of destruction and the quantity of kits that need to be replaced. Customers were also instructed to notify their customers of the advisory notic
Quantity500 kits
DistributionWorldwide distribution to: USA, Mexico, Kuwait, Uraguay, Romania, Italy, Philippines, India, Malaysia, Algeria, Czechoslovakia, Brazil, Colombia, Ireland, Korea, and Saudi Arabia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K901596NOVA LITE(TM) ANA KSL
Inova Diagnostics, Inc., May 2, 1990