Recall Z-0173-2012 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Baxter Healthcare Renal Div

Baxter System 1000 Tina Single Patient Hemodialysis Delivery System; Baxter Healthcare Corporation, Renal Division, McGaw Park, IL 60085; product code S1000L3P. Hemodialysis and Continuous Renal Replacement Therapy devices for acute and chronic hemodialysis

DateNov 7, 2011 (initiated Sep 19, 2011)
StatusTerminated
ReasonThe Tina System was released to the customer with OOS values for the Pressure Displays Accuracy and Leakage Rate Test.
Root causeEmployee error
ActionBaxter Healthcare Corporation sent a service technician to the lone customer site on September 19, 2011 to correct the affected product. The technician executed all tests as established in the Service Manual, with acceptable results.
Quantity1 unit
DistributionDistributed in Puerto Rico only.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K970446DRAKE WILLOCK SYSTEM & ALTRA TOUCH 1000 DIALYSIS DELIVERY SYSTEM
Althin Medical AB an Affiliate of Baxter Intl, Jun 2, 1997