Recall Z-0134-2012 ยท FRN Pump, Infusion

Baxter Healthcare Corp.

Baxter Flo-Gard 6301 Dual Channel Volumetric Infusion Pump; Baxter Healthcare Corporation, Deerfield, IL 60015; Product Code # 2M8064. The Flo-Gard Volumetric Infusion Pumps are an electromechanical pump used for intravenous infusion of fluids at user-selected rates.

DateOct 28, 2011 (initiated Aug 26, 2011)
StatusTerminated
ReasonThe Flo-Gard infusion Pump was released to the customer with OOS values for the Air Sensor Calibrations with the Air in Tubing test.
Root causeEmployee error
ActionA Baxter Healthcare technician initially notified the sole customer on August 26, 2011 of the defective product and requested that the product be quarantined. On September 14, 2011 a technician visited the site to correct the problem observed on the Flo-Gard Infusion Pump. All required tests were performed and the results were found acceptable.
Quantity1 unit
DistributionDistributed only in Puerto Rico.