Recall Z-0272-2012 ยท LLZ System, Image Processing, Radiological

AGFA Corp.

IMPAX CardioVascular (CV) Admin Tool The IMPAX Cardiovascular suite is a cardiovascular information system providing image archiving and image display and modality study-specific structured reporting.

DateNov 23, 2011 (initiated Jun 6, 2011)
StatusTerminated
ReasonPotential for misattribution of data and for a report to become unavailable when the user requests for it to be moved to another study that already has a report associated with it.
Root causeSoftware design
ActionAn 'URGENT FIELD SAFETY NOTICE" was sent via FED-EX by June 6, 2011 to all the sites using the affected product. The letter described the issue and mitigation. The letter recommends that users not move reports to studies where a report already exists. If a report is being moved because the existing report is irrelevant to the study, then the existing report should be deleted prior to moving the new report to that study. If the report is being moved because the user wants to combine the content of the reports, then the "merge" function should be utilized. The letter also states that a soft
Quantity323 units
DistributionWorldwide Distribution -- USA, Virgin Islands and Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050228AGFA HEARTLAB CARDIOVASCULAR
Heartlab, Inc., Feb 18, 2005