Recall Z-0211-2012 ยท LLZ System, Image Processing, Radiological

AGFA Corp.

IMPAX Cardiovascular Results Management (RM) Software Versions RM 2.04.37.04 to RM 7.8 SU2 The IMPAX Cardiovascular suite is a cardiovascular information system providing image archiving and image display and modality study-specific structured reporting.

DateNov 18, 2011 (initiated Jun 6, 2011)
StatusTerminated
ReasonImproper merging of IMPAX Cardiovascular Results Management (RM) reports causes the signed report to become unavailable.
Root causeSoftware design
ActionAGFA Healthcare sent an "URGENT FIELD SAFETY NOTICE" letter dated May 27, 2011, via Fed Ex to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed not to carry out a merge post finalization of an RM reports. If it is necessary to modify a finalized report, AGFA recommends that they make the modification through one of the following actions: Amend the report through the RM interface, or Call AGFA Service to change the status of the report to "preliminary" and then proceed with the finalization of the re
Quantity323
DistributionWorldwide Distribution - USA (nationwide) and Canada
FDA recordOpen on fda.gov