ClearanceBrief
Recall Z-0155-2012 ยท EBF Material, Tooth Shade, Resin

Kerr/pentron Dba Kerr Corporation And Pentron Clinical

37% Etching Gel-1ml syringe 50 pk, part # T06B The intended use of these devices is to etch tooth dentin/enamel in order to prepare the tooth surface prior to the application of a bonding agent.

DateNov 3, 2011 (initiated Mar 31, 2011)
StatusTerminated
ReasonPentron Clinical is voluntarily recalling all lots of products labeled as 37% Phosphoric Acid Etching Gel produced since April 2007 because the product continues to extrude out of the syringe even after pressure is no longer applied, and it contains a higher percentage of phosphoric acid than that indicated on the label.
Root causeNonconforming Material/Component
ActionThe firm, Pentron Clinical, sent an "URGENT: MEDICAL DEVICE RECALL" letter dated May 31, 2011, with attached Return Form (via USPS 1st class mail) to all their consignees/customers. The letter describes the product, problem and action to be taken. All consignees/customers were instructed to complete the Recall Return Form via fax to: 1-877-677-8844 and to return any affected product in their inventory. Affected products will have expiration dates between April 2011 and April 2015. Consignees/customers with questions can contact Pentron Clinical Customer Service at (800) 551-0283 directly to
Quantity41,744 units
DistributionWorldwide distribution: USA (nationwide) and countries of: Armenia, Austria, Australia, Antigua & Barbuda, Barbados, Belarus, Bulgaria, Canada, Czech Republic, Croatia, Cyprus, Ecuador, El Salvador, Finland, France, Germany, Greece, Guatemala, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japa
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K874189DUAL-CURE RADIOPAQUE
Pentron Corp., Feb 11, 1988