Recall Z-0182-2012 ยท LLZ System, Image Processing, Radiological

Philips Medical Systems

Extended Brilliance Workstation-NM (EBW-NM) with software version 1.0P, 1.1.1A, 1.5H, 1.5.1A and 2.0Q -Field Test Period A nuclear medicine image display and processing application suite that provided software applications used to process analyze and display medical images/data.

DateNov 10, 2011 (initiated Sep 15, 2011)
StatusTerminated
ReasonA problem related to a software error has been detected in the Philips Multiple Gated Radionuclide Angiocardiography (MUGA) application. A false negative interpretation due to a falsely elevated ejection fraction or a false positive interpretation due to a false low ejection fraction could occur.
Root causeSoftware design
ActionPhilips sent all affected sites an Urgent Medical Device Correction Letter on September 27, 2011 for the MUGA issue. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review the information contained in the letter with all members of their staff who need to be aware of the contents of the communication. Customers were asked to retain a copy of the communication with the equipment instruction for use. Additionally, a field change order was released on September 26, 2011, customers will be receiving release notes as part
Quantity1181 systems
DistributionWorldwide Distribution - USA (nationwide) and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K012009MX8000 IDT CT SYSTEM
Marconi Medical Systems, Inc., Sep 13, 2001
K080961NM APPLICATION SUITE
Philips Nuclear Medicine, Inc., Apr 18, 2008