Recall Z-0151-2012 ยท LXH Orthopedic Manual Surgical Instrument

Encore Medical, Lp

RSP Baseplate Rim Planer (Size 32, 36, 40): DJO Surgical Instruments and Instrument Cases DJO surgical Encore Medical, L.P. 9800 Metric Blvd, Austin, TX 78758 Contact information: DJO Surgical, ATTN: Customer Service, 9800 Metric Blvd, Austin, TX 78758, 1-800-456-8696 The Baseplate Rim Planer is used to prepare the bone around the peripheral rim of the reamed socket. It has a spring loaded alig

DateNov 2, 2011 (initiated Sep 28, 2011)
StatusTerminated
ReasonDJO Surgical has determined that the alignment pins cannot be consistently cleaned using the technique provided in the cleaning/sterilization section of the instrument IFU.
Root causeDevice Design
ActionDJO Surgical sent an URGENT FIELD SAFETY NOTICE letter dated September 28, 2011, to all affected customers. The letter identified the product. the problem, and the action to be taken by the customer. Customers were instructed to return the RSP Humeral Socket Reamers and RSP Baseplate Rim Planers. The RSP Humeral Socket Reamers will be reworked and returned to the customer. The Baseplast Rim Planer will be replaced when a suitable design is available. During the interim until instruments are returned, the use of FMP acetabular reamers size 42mm, 43mm, 45mm (DJO PIN 803-05-226, 227, 200)
Quantity133 Planers
DistributionWorldwide Distribution - USA including AR, AZ, CA, CO, FL,GA, ID, IN, IL, LA, MA, MD, ME, MI, MO, NJ, NM, NY, OH, OK, PA, PR, SC, TN, TX, UT, VA, WA, and WI and Internationally toGermany, South America, Italy, Saudi Arabia, United Kingdom, and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.