ClearanceBrief
Recall Z-0255-2012 ยท KPS System, Tomography, Computed, Emission

Philips Medical Systems

Precedence System; SYST,PRECEDENCE, 16 SLICE 5/8 Manufactured by Philips Medical Systems (Cleveland), Inc. 595 Miner Road, Cleveland, OH 44143. The SKYLight Gamma Camera System is intended to product images depicting the anatomical distributions of single photon and positron emitting radioisotopes within the human body for interpretation by medical personnel.

DateNov 19, 2011 (initiated Sep 27, 2011)
StatusTerminated
ReasonA mechanical malfunction of the button on the hand controller on the Philips Precedence and SKYLight SPECT systems may get stuck and cause unintended motion of a detector.
Root causeDevice Design
ActionPhilips Medical Systems sent a "Customer Information" letter dated September 27, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The letter lists additional precautions the customers should use while operating the device. Philips will notify customers for replacing the affected product through an Action for Performance-Proactive Field Change Order to resolve the issue. Contact Philips Customer Care Solutions at 1-800-722-9377, option 5, for questions concerning this notice.
Quantity658 - all units in recall
DistributionWorldwide Distribution-USA (nationwide) including Puerto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041218GRIFFIN SPECT/CT IMAGING SYSTEM
Adac Laboratories, May 24, 2004