Recall Z-0077-2012 ยท DXS Gauge, Pressure, Coronary, Cardiopulmonary Bypass

Medtronic Perfusion Systems

Medtronic Pressure Display Box, Models 66000, 64000, 63000, 60000. It is capable of measuring pressure between -100 mmHg and +500 mmHg with an accuracy of +/-5 mmHg from -100 mmHg to +300 mmHg and +/-10 mmHg from +301 to +500 mmHg. The Pressure Display Box has an upper and lower alarm setting. The alarm will be activated when the pressure is equal to or beyond the set value. Manufacturer: Me

DateOct 20, 2011 (initiated Sep 9, 2011)
StatusTerminated
ReasonThe DFU for the Medtronic Pressure Display Box, model 66000, states that, for display readings of -100 to +300 mmHg there will be an accuracy of +/-5 mmHg, and for display readings of +301 to +500 mmHg there will be an accuracy of +/-10 mmHg. Through reports from customers, Medtronic has become aware that the Pressure Display Box may not always accurately display readings. During use in the -40
Root causeDevice Design
ActionMedtronic Perfusion Systems, sent an "Urgent Medical Device Notification letter dated September 9, 2011, via UPS 2-day delivery to all affected customers. The letter identified the affected devices, described the issues, and the actions needed to be taken. The letter indicates when approved, the updated Directions for Use for the Pressure Display Box and Pressure Display Sets will be sent to customers. Customers were asked to complete the Customer Confirmation Certificate and return to Medtronic per the instructions on the form. The letter states that if the recalled products have been fu
Quantity1694 units
DistributionWorldwide Distribution - USA (nationwide) including Puerto Rico, and Virgin Islands (U.S.) and the countries of: Argentina, Australia, Austria, Belarus, Belgium, Bulgaria, Canada, Chile, China, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, El Salvador, France, Georgia, Germany, Greece, Hon
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K852232PRESSURE DISPLAY BOX #60000/TUBING SET #61000
Dlp, Inc., Aug 9, 1985