Recall Z-0282-2012 ยท JAQ System, Applicator, Radionuclide, Remote-Controlled

Varian Medical Systems, Inc.

Varian Medical Systems, Flexible Probe with Blocking Water, GM11002420, (Part of the Segmented Cylinder Set, GM11004150)

DateNov 23, 2011 (initiated May 20, 2010)
StatusTerminated
ReasonThe Instructions For Use ( IFU) distributed with product prior to 2004 lacks the correct limit to the sterilization cycles and checks regarding the firm seating of the blocking washer.
Root causeLabeling Change Control
ActionVarian Medical Systems sent a Medical Device Recall (PNL-GM-42697-2) in May 2010 for the blocking washer on the flexible probe, Varian part number GM11002420, a component of the Segmented Cylinder Applicator Set. The letter was to inform users that a flexible probe with an unstable or shifted blocking washer must not be used due to the potential for treatment length error. The blocking washer may move from rough handling or sterilization. Users were told to discard any instances of the obsolete Instructions for Use (IFU) in use and replace it with the IFU (PGM11004150 4 March 24, 2010.pdf).
Quantity81 parts
DistributionWorldwide Distribution: USA, Germany, Belgium, Canada, Mexico City, Spain, Brazil, Switzerland, Uzbekistan, India
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K033371VARIAN INTRACAVITARY BRACHYTHERAPY APPLICATORS
Varian Medical Systems, Dec 24, 2003