Recall Z-3275-2011 ยท EPT Microscope, Surgical

Carl Zeiss Meditec, Inc.

OPMI Pentero, OPMI Pentero C, OPMI Pentero 900, OPMI Vario/S8, OPMI Vario/S88, OPMI Neuro Multivision/NC4, OPMI Vario/NC33, OPMI Neuro/NC4, OPMI Neuro/NC4 Ceiling Mount. A surgical microscope and accessories is an AC powered device intended for use during surgery to provide a magnified view of the surgical area.

DateSep 23, 2011 (initiated Sep 12, 2011)
StatusTerminated
ReasonSome third party surgical drapes have optically active protective lenses that could affect the ZEISS surgical microscope position information that is passed on to third party navigation systems.
Root causePending
ActionThe firm, Zeiss, sent an "IMPORTANT SAFETY ALERT FOR ZEISS SURGICAL MICROSCOPES" letter dated September 7, 2011, to all affected customers. The letter described the product, problem and action to be taken. The customers were instructed to use surgical drapes compatible with ZEISS surgical microscopes and always follow the instructions for use of the third party navigation system manufacturer for calibration, calibration verification and operation. In addition, the firm supplied customers with a supplement to their Zeiss surgical microscope user documentation and requested that they read the
Quantity1677 consignees
DistributionNationwide distribution: USA only.
FDA recordOpen on fda.gov