Recall Z-0069-2012 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Siemens Healthcare Diagnostics, Inc.

Siemens Healthcare Diagnostics, Inc. Dimension(R) EXL Cardiac Troponin I Flex(R) Reagent Cartridges

DateOct 19, 2011 (initiated Aug 31, 2011)
StatusTerminated
ReasonFirm has confirmed customer complaints for low recovery of QC and patient samples.
Root causeProcess control
ActionSiemens Healthcare Diagnostics, Inc. Urgent Field Safety Notice was sent to customers on August 31, 2011 regarding the issue. The letter instructs customers to immediately discontinue use of the affected lots and discard all remaining inventory.
Quantity6,072 cartons
DistributionWorldwide distribution: USA, Puerto Rico, Guam, Austria, Canada, Croatia, France, Greece, Germany, Italy, Japan, Mexico, Portugal, Spain and Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081643DIMENSION TNI FLEX REAGENT CARTRIDGE, AND CTNI SAMPLE DILUENT WITH MODELS, RF621, KD692