Recall Z-0325-2012 ยท JAQ System, Applicator, Radionuclide, Remote-Controlled
Varian Medical Systems, Inc.
VariSource iX, Software Version 1.1, Part Number AL20020000 VariSource iX Afterloader. The VariSource iX Series High Dose Rate Afterloader system is a computer controlled remote electra/mechanical system used for medical purposes, for placing a NiTiNol wire incorporating an irradiated iridium seed internally or close by, a malignant tumor or tumor bed in a practice known as brachytherapy. The dev
| Date | Dec 1, 2011 (initiated Mar 1, 2010) |
| Status | Terminated |
| Reason | VariSource iX console software may cause incorrect dwell times to be delivered to patient and may result in a mistreatment. |
| Root cause | Software design |
| Action | Varian Medical Systems sent a "MEDICAL DEVICE RECALL" letter dated April 1, 2010 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. Customers were asked to conduct a simple test to establish if their machine is affected by the problem. Device users were requested to complete an attached receipt verification form and return it to Varian BrachyTherapy via fax or e-mail. Call the Contact Service Dispatch number at 1-800-864-1672 for questions regarding this detail. |
| Quantity | 72 worldwide/50 in U.S. |
| Distribution | Worldwide Distribution-USA (nationwide) |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K071467 | VARISOURCE IX HDR AFTERLOADER Varian Medical Systems, Inc., Jul 16, 2007 |