Sodium Bicarbonate Mixing and Delivery System Software v 7.1. Model # MD-380, Serial Numbers: 5103031102, 5101101101, 5102161007, 5103221007, 5104221007, 5105101008, 5109131010, 5111231011, 5114301012, 5102131101 & 5113091012. Manufactured by: Isopure Corp, Simpsonville, KY System is used to mix and distribute bicarbonate as a part of a clinical dialysis system.

DateOct 11, 2011 (initiated May 19, 2011)
StatusTerminated
ReasonSoftware problem causes transfer line between mixing and storage tank to not be rinsed after disinfection, leaving residual chlorine in the line.
Root causeSoftware design
ActionIsopure sent a letter dated May 19, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Isopure devised an immediate solution which must be implemented at their faclity in conjunction with the disinfection procedure. Customers were asked to inform all appropriate personnel, clinical and technical, who may be involved in operating the system during the normal disinfection schedule. On 5/19/2011, the firm sent letters to their customers via US mail and email. They also telephoned their customers to notify them of the p
Quantity11 units
DistributionNationwide Distribution including AR, CT, GA, MA, ME, NJ, PA, TN and WV.