Recall Z-3270-2011 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories
Maquet Cardiovascular, LLC
VASOVIEW 6 Endoscopic Vessel Harvesting System Manufactured by Maquet Cardiovascular, LLC; 170 Baytech Drive, San Jose, CA 95134. Product Usage: Endoscopic vessel harvesting systems are indicated for use in minimally invasive surgery allowing access for vessel harvesting and are primarily indicated for patients undergoing endoscopic surgery for arterial bypass.
| Date | Sep 20, 2011 (initiated Dec 17, 2009) |
| Status | Terminated |
| Reason | During routine production quality monitoring process, it was found that UV curing adhesive was not used in a limited number of devices. |
| Root cause | Nonconforming Material/Component |
| Action | MAQUET Cardiovascular, LLC, sent an "Urgent Device Removal/Immediate Action Required" letter dated December 17, 2009, to all affected customers. The letter identified the recalled products, the problem, and the actions needed to be taken. The letters asked customers to examine their stocks (inventory) immediately to determine if they had any affected lot numbers and to discontinue dispensing (distributing) the lot. The letter states that customers are to complete the attached Field Action Response form regardless of whether or not the affected products are located in their inventory. Distri |
| Quantity | 141 units |
| Distribution | Nationwide Distribution (USA) including the states of: AL, AZ, CA, CT, FL, KS, KY, IL, IN, MI, MO, ND, NY, NV, OH, OK, OR, PA, SC, TN, TX, UT, WA, and WI. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K041981 | GUIDANT VASOVIEW 6 HARVESTING CANNULA, MODEL VH-2000; GUIDANT VASOVIEW 6 ACCESSORY PACK, MODEL VH-2004 Guidant Corporation, Aug 20, 2004 |