Recall Z-0051-2012 ยท FRN Pump, Infusion

B Braun Medical, Inc

Outlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used to regulate the flow of primary and secondary fluids when positive pressure is required. The infusion system is capable of delivering fluid from a negative head height (when the IV fluid container is lower than the pump), and provides clinically accepted volumetric accuracy for all standard IV fluids, includi

DateOct 18, 2011 (initiated Sep 7, 2011)
StatusTerminated
ReasonB Braun Outlook ES Safety Infusion System "Run" Light Emitting Diodes (LEDs) pumps stop infusing and a backup alarm sounds, but the "Run" LEDs advance as if the pumps were infusing.
Root causeSoftware design
ActionB. Braun Medical, Inc. sent an "URGENT: FIELD CORRECTION" letter dated September 7, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. A firm representative will contact all affected customers in order to schedule the field correction. The correction consists of revising the Door and Main Processor software, inspecting the main board, and replacing the board, if warranted. A Field Correction Information Form was included in the letter for customers to complete and return. Contact Technical Product Support at 1-800-627-7867
Quantity6321
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K011975HORIZON OUTLOOK WITH DOSECOM
B.Braun Medical, Inc., Sep 19, 2001
K994375HORIZON LIGHT
B.Braun Medical, Inc., Mar 8, 2000