ClearanceBrief
Recall Z-0586-2012 ยท MSD Catheter, Hemodialysis, Implanted

Angiodynamics, Inc.

DURAMAX 15.5 F x 22 cm Chronic Hemodialysis Catheter Set (With Cuff 17 cm from Tip), DuraMax Stacked Tip 22 cm Str. Basic Kit, Catalog No./REF 10302802, STERILE --- AngioDynamics, Inc. 603 Queensbury, NY USA 12804. The AngioDynamics DuraMax Chronic Hemodialysis Catheter is indicated for use in attaining long-term vascular access for Hemodialysis and apheresis.

DateJan 11, 2012 (initiated Jan 24, 2011)
StatusTerminated
ReasonAngioDynamics received customer complaints for breakage of the tunneler sleeve during product use.
Root causeNonconforming Material/Component
ActionThe firm, AngioDynamics, sent an "URGENT: MEDICAL DEVICE RECALL" letter dated January 28, 2011 and response form via Certified Mail, Return Receipt Requested, to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to identify and segregate the recalled lots; complete and return the enclosed DuraMax HemoDialysis Catheter Recall Reply Form via fax to: Attn: DuraMax HemoDialysis Catheter Recall Coordinator at 518-798-1360 (even if they don't have any product); and do not use the affected product, return and ship the recalled product to
Quantity12 units
DistributionNationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K080400ANGIODYNAMICS, INC., DURAMAX HEMODIALYSIS CATHETER AND PROCEDURE KIT
AngioDynamics, Inc., May 13, 2008
K101843DURAMAX CHRONIC HEMODIALYSIS CATHETER AND PROCEDURE KIT
AngioDynamics, Inc., Oct 20, 2010