Recall Z-1041-2012 ยท CBK Ventilator, Continuous, Facility Use

Flight Medical Ltd.

V24-00400-29 cable, an accessory to the Newport HT50 ventilator. The V24-00400-29 cable is an accessory to the Newport HT50 ventilator intended to transfer the alarm signal from the ventilator to a nurse remote station.

DateFeb 21, 2012 (initiated Aug 10, 2011)
StatusTerminated
ReasonThe recall was initiated because Flight Medical Innovations Ltd. has identified a problem may occur when contact between an exposed unused wire and the grounding wire occurs. If this contact occurs the ventilator may reset and switch off.
Root causeDevice Design
ActionNewport Medical Instruments, Inc., the U.S. distributor for Flight Medical Ltd., sent a letter dated August 23, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. The letter instructs customers to immediately discontinue use of the affected product. An Acknowledgement Form was attached to the letter for customers to complete and return via fax at (714) 427-0489 or e-mail to Regulatory@NewportMedical.com. Replacement product will be provided once the form is received. Contact Newport Medical's Technical Service Group at 800-
Quantity363 total units, all in the US
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082724NEWPORT HTS50 VENTILATOR WITH DUAL PAC BATTERY SYSTEM
Newport Medical Instruments, Inc., Dec 16, 2008
K992133NEWPORT HT50 VENTILATOR WITH HUMIDIFIER HT50-H, NEWPORT HT50 VENTILATOR WITHOUT HUMIDIFIER, HT50-H1, NEWPORT HT50 VENTIL
Newport Medical Instruments, Inc., Aug 4, 2000