Recall Z-0210-2012 ยท LYJ Stimulator, Autonomic Nerve, Implanted For Epilepsy
Cyberonics, Inc
VNS Therapy Aspire HC Generator and VNS, Model 105 Cyberonics, Inc. 100 Cyberonics Blvd., Houston, TX 77058 For implantation in humans to treat Epilepsy and Depression.
| Date | Nov 18, 2011 (initiated Aug 12, 2011) |
| Status | Terminated |
| Reason | The devices are being recalled because the output current delivered to the vagus nerve is less than the design intent and there is a potential charge imbalance at the lead cathode and generator-can during stimulation. |
| Root cause | Nonconforming Material/Component |
| Action | Cyberonics sent a Field Safety Alert on August 15, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. The firm decided to recall and initiated contact with implanting facilities in the United States on 08/15/2011. A Field Safety Alert was delivered on 09/19/2011 to implanting facilities. Customers were instructed not to prescribe AspireHC pulse generators for implant until Cyberonics resumes shipment of AspireHC pulse generators that have the issue remedied. Cyberonics recommends that physicians continue their cur |
| Quantity | 340 units |
| Distribution | Worldwide Distribution - USA including TN, TX, IL, OH, FL, CA, CO, OR, AZ, NY, PA, MO, NJ, WI, OK, MI, ID, KY, IA, MA, NE, GA, LA, NY, and Norway. |
| FDA record | Open on fda.gov |