Recall Z-3276-2011 ยท LHI Set, I.V. Fluid Transfer

Baxter Healthcare Corp.

LOGIX Order Entry (OE) Software, all versions, catalog 2M8410; Baxter Healthcare, Deerfield, IL 60015; The software is included in a number of Baxter's pharmacy compounders. The software assists the pharmacist in management of compounding, creating and authorizing nutritional solutions intended for IV administration. Product Usage: The LOGIX OE application is an order entry system intended for

DateSep 23, 2011 (initiated Aug 15, 2011)
StatusTerminated
ReasonThere is the potential for an order entry error during use of the LOGIX Order Entry (OE) software. When adding ingredients to a TPN order, the user can accept the pre-populated "per units" of "Bag" or select other "per units". A dosing error may occur if the user does not select the prescribed "per units" for the added ingredient.
Root causeSoftware design
ActionBaxter Healthcare Corp. sent an "Urgent Device Correction" letters dated August 15, 2011 which included LOGIX OE screen illustrations of a representative neonate and adult patient, via first class mail to all affected customers. The letter identified the product, problem and actions to be taken. The customers were instructed to follow safe compounding practice, which includes verifying that the ingredient per units entered match the TPN prescription prior to preparation. Baxter requests that customers communicate this issue to all personnel responsible for LOGIX order entry and order author
Quantity95 units
DistributionWorldwide Distribution - USA (nationwide) including Puerto Rico and the country of Canada.
FDA recordOpen on fda.gov