Recall Z-3263-2011 ยท GGK Products, Red-Cell Lysing Products

Beckman Coulter Inc.

COULTER DxH Cell Lyse, Part number 628018(1L) and 628019 (5L). The COULTER DxH Cell Lyse is an erythrocyte lytic agent used for quantitative determination of hemoglobin, enumeration of NRBC, and counting and sizing leukocytes on UniCel DxH 800 Coulter Cellular Analysis System.

DateSep 19, 2011 (initiated Jul 11, 2011)
StatusTerminated
ReasonThe recall was initiated because Beckman Coulter identified unexpected high particle counts in the lots of DxH Cell Lyse listed. These lots may generate high background counts; could result in a falsely elevated Total Nucleated Count (TNC) of up to 50 cells/L; or may also result in recovery of TNC values that were higher than expected for Level 1 of the Body Fluid Control.
Root causePending
ActionThe recall communication was initiated on 08/01/2011 with Beckman Coulter forwarding a Product Corrective Action (PCA) letter with attached PCA Response Form (via US Postal Service for US customers) to all customers, who purchased Coulter DxH Cell Lyse. The letter provides the customers with an explanation of the problem identified and an action to be taken. Customers were instructed to contact Beckman Coulter for product replacement. In addition, customers were instructed to complete and return the enclosed PCA Response Form within 10 days. Customers with questions regarding this Product
Quantity4913 units total (2245 in US)
DistributionWorldwide Distribution -- Australia, Belgium, Brunei Darussalam, Canada, France, Germany, Hong Kong, Hungary, India, Italy, Japan, Macao, Malaysia, Netherlands, Norway, Philippines, Qatar, Russian Federation, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Un
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.