Recall Z-3046-2011 ยท JOJ Indicator, Physical/Chemical Sterilization Process

Steris Corporation

Verify SixCess Challenge Pack, Model #'s: LCC003, LCC005, LCC014 & LCC310, Mfr. By: Albert Browne, LTD, Chancery House 190 Waterside Rd. Hamilton Industrial Park, Leicester, Leicester, LE5QZ, United Kingdom. Verify SixCess Challenge Packs are used to confirm that that critical parameters of steam sterilization have been reached within a challenging test pack design. Each challenge pack is also s

DateAug 22, 2011 (initiated Mar 16, 2011)
StatusTerminated
ReasonDuring an inspection the firm was notified that language contained in labeling and promotional literature directs Customers that the products can be used to replace biological indicators and/or release loads containing implants.
Root causeLabeling False and Misleading
ActionSTERIS sent a Customer Notification letter dated March 16, 2010, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review and follow the Challenge Packs' revised instructions for use enclosed as Attachment A (270-4) and Attachment B (275F 3). For any questions regarding this recall customers were instructed to contact their STERIS Account Manager or call 440-392-7421.
Quantity2,500
DistributionWorldwide Distribution - USA (nationwide) and the countries of Lebanon and Qatar.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.