ClearanceBrief
Recall Z-3197-2011 ยท JAK System, X-Ray, Tomography, Computed

Toshiba American Medical Systems Inc

TSX-201A: AQUlLlON LB; System, X-Ray, Tomography, Computed X-Ray Tomography

DateSep 13, 2011 (initiated Jun 22, 2011)
StatusTerminated
ReasonToshiba America Medical Systems is initiating a field correction on their TSX-301B: AQUILION Premium; System, X-Ray, Tomography, Computed, TSX-301A: AQUILION One; System, X-Ray, Tomography, Computed, TSX-20IA: AQUlLlON LB; System, X-Ray, Tomography, Computed, and TSX-1O1A: AQUILION 16; System, X-Ray, Tomography, Computed due to several software issues that cause the scanners to function incorr
Root causeSoftware design
ActionToshiba America Medical Systems (TAMS) sent an URGENT MEDICAL DEVICE CORRECTION letter dated June 22, 2011, to all affected customers. Customers were instructed that a revised software to prevent occurrence of the problem would be installed on their system. Their Toshiba Service Representative would contact them for an appointment to install the software to their system when it became available. Customers were requested to take temporary measures until corrective measures were taken. 1) In dual scanoscopy, do not press the short key between the completion of the first scanoscopy and the st
Quantity155 units total
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050458TSX-201A AQUILION LB CT SCANNER
Toshiba America Medical Systems, In.C, Mar 10, 2005