ClearanceBrief
Recall Z-3040-2011 ยท LGZ Warmer, Thermal, Infusion Fluid

Vital Signs Devices, a GE Healthcare Company

enFlow IV Fluid Warmer Part number 980105VS; (510K name: enFlow IV Fluid Warmer Model 100, EFlow Disposable Cartridge, Model 200 and EFlow AC Power Supply Model 120) Intended for warming blood, blood products, and intravenous solutions, prior to administration. Intended to be used by healthcare professionals in hospital, clinical and field environments to help prevent hypothermia.

DateAug 19, 2011 (initiated Jun 30, 2011)
StatusTerminated
ReasonenFlow IV Fluid Warmer may have an issue which results in excessive external surface temperature on the underside of the warmer.
Root causeDevice Design
ActionGE Healthcare sent UMDC Letters/UMDC Confirmation Letters to customers on June 30, 2011. GE Healthcare sent UMDC Letters/UMDC Confirmation Letters on October 24, 2011 reflecting a new recall strategy.
Quantity1307 units
DistributionWorldwide Distribution - USA (nationwide) and the country of Vietnam
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060537EFLOW IV FLUID WARMER, MODEL 100, EFLOW DISPOSABLE CARTRIDGE, MODEL 200 AND EFLOW AC POWER SUPPLY, MODEL 120
Enginivity, LLC, Jun 8, 2006