Recall Z-3220-2011 ยท NQX System, Nucleic Acid Amplification Test, Dna, Methicillin Resistant Staphylococcus Aureus, Direct Specimen
Becton Dickinson & Co.
BD MAX GBS Catalog Number 441772 For In Vitro Diagnostic Use, For use with the BD MAX system.The Sample Preparation Reagent kit consists of 24 (2 x 12) 2D barcode tubes which verifies against the number entered in the worklist with the handheld barcode reader. The BD MAX System is a qualitative in vitro diagnostic test obtained from vaginal and rectal swab specimens from antepartum pregnant women
| Date | Sep 16, 2011 (initiated May 24, 2011) |
| Status | Terminated |
| Reason | Medical device bar code reader does not function as intended and could result in associating clinical test results to wrong patient. |
| Root cause | Nonconforming Material/Component |
| Action | BD Diagnostic Systems sent an "URGENT PRODUCT RECALL" letter dated May 2011, to all affected customers. The letter identified the products, the problem, and the action to be taken by the customers. The letter advised customers to discontinue use of the Sample Preparation Reagent tubes in the referenced kits and discard any remaining products. Customers were requested to complete an attached reply form to acknowledge their receipt of the notification and fax it to (410) 316-4258. For further assistance regarding replacements contact BD Customer Service Dept. 1-800-675-0908. For all other i |
| Quantity | 61 kits |
| Distribution | Worldwide Distribution-USA (nationwide) including the states of CA, KY, MD, MI, NC, NY, TX, and WI and the country Belgium. |
| FDA record | Open on fda.gov |