Recall Z-3218-2011 ยท NQX System, Nucleic Acid Amplification Test, Dna, Methicillin Resistant Staphylococcus Aureus, Direct Specimen

Becton Dickinson & Co.

BD MAX DNA Extraction Kit DNA-3 (Swabs in Transport Medium/UTM), Catalog Number 437503. Kit contains 24/2D barcode sample preparation reagent tubes ***For use with the BD MAX System *** to extract DNA from Gram-negative bacteria and viruses, which may be present in clinical specimens. The specimen is mixed with BD MAX Sample Preparation Reagent and processed using the BD MAX System. The BD MA

DateSep 16, 2011 (initiated May 24, 2011)
StatusTerminated
ReasonSample results maybe switched between time it was scanned with handheld barcode reader and time it was placed in the sample rack, associating sample results to wrong patient.
Root causeNonconforming Material/Component
ActionBD Diagnostic Systems sent an "URGENT PRODUCT RECALL" letter dated May 2011, to all affected customers. The letter identified the products, the problem, and the action to be taken by the customers. The letter advised customers to discontinue use of the Sample Preparation Reagent tubes in the referenced kits and discard any remaining products. Customers were requested to complete an attached reply form to acknowledge their receipt of the notification and fax it to (410) 316-4258. For further assistance regarding replacements contact BD Customer Service Dept. 1-800-675-0908. For all other i
Quantity163 kits
DistributionWorldwide Distribution-USA (nationwide) including the states of CA, KY, MD, MI, NC, NY, TX, and WI and the country Belgium.