Recall Z-3047-2011 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Draeger Medical Systems, Inc.

Infinity Delta Portable Patient Monitor. Draeger Medical Systems, Inc. Portable monitor used to monitor vital signs of neonatal, pediatric, and adult patients for bedside and transport applications.

DateAug 22, 2011 (initiated Jul 29, 2011)
StatusTerminated
ReasonMonitor keys may spontaneously become inoperative or active. This may cause a membrane switch panel malfunction, which could result in the monitor discharging a patient automatically.
Root causeDevice Design
ActionDraeger Medical sent an "Urgent Medical Device Recall Notification" letters dated July 2011 to all affected customers. The letter included product and problem affected. The letter informed the customer a Draeger representative will contact them to upgrade the system. Until the solution is implemented, Draeger recommends the Delta and/or Gamma XXL may be used. If the affected monitors become inoperative or activate without user interaction, Draeger advises removing the monitor from service and contacting a Draeger representative. For questions or for information on this recall please cal
Quantity32
DistributionWorldwide Distribution - (USA) Nationwide distribution including the states of AK, FL, LA, MN, MS, MO, NJ, NC, OH, OK, PA, TN, TX, NC, NH, NJ, NY, WV, and WI; and the countries of Australia, Austria, Argentina, Bahrain, Brazil, Bulgaria, Canada, China, Colombia, Czech Republic, France, Germany, Geor
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K070566INFINITY DELTA/DELTA XL/ KAPPA/ GAMMA X XL/ VISTA XL/ SC7000/SC8000/ SC9000XL
Draeger Medical Systems, Inc., Jun 4, 2007