Recall Z-3214-2011 ยท DWB Pump, Blood, Cardiopulmonary Bypass, Roller Type

Terumo Cardiovascular Systems Corporation

Sarns 8000 Cardioplegia Monitor, Catalog Number 16414; also known as PROG COMP BD AMPRO 420 CPG 8K, Catalog Number 816373 The Sarns 8000 Cardioplegia Monitor is indicated to measure extracorporeal line pressure and signal when the pressure exceeds the operator-set limit.

DateSep 15, 2011 (initiated Aug 29, 2011)
StatusTerminated
ReasonThe Cardioplegia Monitor for the Sarns Modular Perfusion System 8000 may fail to track cardioplegia delivery volume.
Root causeComponent change control
ActionTerumo Cardiovascular Systems sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated August 29, 2011 to all affected customers. The letter describes the product, problem, and action to be taken. The letter advises customers to make sure all users are aware of the notice. Customers are instructed to complete and return an attached Customer Response Form. Terumo will replace the software on all monitors of the affected product. Customers with questions or concerns should call Terumo CVS Customer Service 1-800-521-2818.
Quantity85
DistributionWorldwide Distribution--USA (nationwide) including the states of AL, AR, CA, CO, GA, IL, IN, MS, NC, OH, TN, VA, and WI and countries of Belgium, Columbia, Hong Kong, India, Singapore, and Taiwan.