Recall Z-3282-2011 ยท LLZ System, Image Processing, Radiological

GE Healthcare, LLC

ADVANTAGE WINDOWS (Workstation) WITH FUNCTOOL OPTION GE Medical Systems, SCS. A software module that provides supplemental information to those images extracted from CT and MR temporal datasets.

DateSep 27, 2011 (initiated Aug 18, 2011)
StatusTerminated
ReasonGE Healthcare has become aware that image orientation of reference series displayed in Functool Diffusion Tensor/FiberTrak protocol on Advantage Workstation may not match with the orientation annotations displayed in the viewport. This issue may impact patient safety.
Root causeSoftware design
ActionGE Healthcare sent "Urgent Medical Device Correction" letters dated August 18, 2011 to all affected customers. The letter included name of product affected and described problem. Recommended actions where provided to customers and they were notified a GE Healthcare representative will contact them for a software system upgrade. For question of this recall contact the GE Service Representative at (800) 437-1171.
Quantity697
DistributionWorldwide Distribution: Nationwide distribution including the states of AZ, CA, CT, DC, FL, ID, IL, KY, LA, MD, MI, MN, NJ, NY, OR, PA, PR, TN, TX, VA, WA, and WI; and the countries of VIETNAM, MEXICO, UNITED KINGDOM, TURKEY, TUNISIA, THAILAND, TAIWAN, SWITZERLAND, SWEDEN, ST HELENA, SPAIN, SLOVEN
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K960265ADVANTAGE WINDOWS WITH FUNCTOOL OPTION
GE Medical Systems, Jul 3, 1996