Recall Z-3030-2011 ยท JPD Hemoglobin A2 Quantitation

Tosoh Bioscience Inc

Hemoglobin F & A2 Calibrator, 0.2 mL, 4 brown glass vials/box. Mfr. By: Tosoh Bioscience, Minato-ku, Tokyo, Japan. Hemoglobin F & A2 Calibrator is used as in-vitro calibration material for Beta-thalassemia testing on the Tosoh G7.

DateAug 17, 2011 (initiated May 25, 2011)
StatusTerminated
ReasonThe value after reconstitution appeared to be lower than the assigned value on the label. The measured value for Lot BCAL05R was 5.3%, whereas the value on the label is 5.7%. The criteria of the value is +/- 0.3%.
Root causeProcess control
ActionTOSOH Bioscience sent an "Urgent Recall Notification" letter dated July, 25, 2011 to all affected customers. The letter described the problem and asks customers to immediately discontinue use of the product and return all remaining product back to the firm. Customers can contact TOSOH for replacement. For additional information or question on this recall please contact TOSOH Technical Support at (800) 248-6769.
Quantity6 boxes/4/2 mL btls
DistributionWorldwide Distribution: Nationwide (USA) distribution including the states of FL and IL; and the country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K020489G7 AUTOMATED HPLC ANALYZER, BETA-THALASSEMIA MODE
Tosoh Medics, Inc., May 14, 2002