Recall Z-3083-2011 ยท ONW Hemodialysis System For Home Use

Fresenius Medical Care Holdings, Inc.

Fresenius 2008K@home The Fresenius 2008K@home is indicated for acute and chronic dialysis therapy in an acute or chronic facility. The 2008K@home is also indicated for hemodialysis in the home and must be observed by a trained and qualified person as prescribed by their physician.

DateAug 29, 2011 (initiated May 4, 2011)
StatusTerminated
ReasonUpgrade software to implement new on screen graphics and instructions, the addition of a bracket on the side of the machine to temporarily hold the bloodlines during set up in order to conform to approved 510(k)
Root causeDevice Design
ActionFresenius Medical NA issued a formal field correction notification dated May 4 ,2011, sent by certified mail with return receipt requested. The letter identified the product the problem and the action needed to be taken by the customer. The customer was notified of the retrofit that would be done to their machines. The patients' facilities were made aware of the changes that would be made to each of their patient's 2008K@home machines on May 4, 2011. On May 13, 2011, all facilities with active 2008K@home machines were sent a notification letter of the retrofit. Please return the receipt
Quantity705 units
DistributionNationwide Distribution. Canada - FMCNA is not conducting the recall to Canada, as Health Canada did not require the same modifications to the machines
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K070049FRESENIUS 2008K @HOME WITH WIRELESS WETNESS DETECTOR
Fresenius Medical Care North America, Feb 3, 2011