Recall Z-3205-2011 ยท LLZ System, Image Processing, Radiological

Philips Healthcare Inc.

PCR Eleva Radiological Image Processing System with software R.1.2 Philips Healthcare, Inc. A computed radiography image processing system.

DateSep 14, 2011 (initiated Jul 27, 2011)
StatusTerminated
ReasonThe mirror icon on the bottom of each image could be interpreted as a RIGHT marker for side indication if the image is mirrored again within a PACS system. This could potentially lead to a misdiagnosis
Root causeSoftware design
ActionPhilips Healthcare Inc. sent out "Urgent --Medical Device Correction" notices dated June 10, 2011 to all affected customers. The letter described the product and problem. Customers were informed the "burned-in Mirror-Icon" may be misinterpreted as a Right marker. Customers are advised to preferably use lead markers during each X-ray acquisition or alternatively electronic annotations on each image for side indication. The presentation of these markers within each image will clearly show the difference between the intended Left/Right indication and the little mirror ICON. Philips will ins
Quantity197 units
DistributionWorldwide Distribution: Nationwide distribution; and including the countries of Albania, Angola, Argentina, Australia, Austria, Azerbaijan, Belgium, Bolivia, Brazil, Bulgaria, Canada, Chile, China, Croatia, Denmark, Egypt, France, Germany, Ghana, Greece, Greece, Haiti, Hungary, Indonesia, Iraq, Isr
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K964124PHILIPS COMPUTED RADIOGRAPHY
Philips Medical Systems, Inc., Dec 18, 1996