ClearanceBrief
Recall Z-3037-2011 ยท DRF Catheter, Electrode Recording, Or Probe, Electrode Recording

Biosense Webster, Inc.

ISMUS Catheter Mapping Catheter with Auto ID Technology, Manufacturing Part Number D-1171-34-S, Catalog Number D7R20P14CT The ISMUS Catheter with Auto ID Technology has been designed for electrophysiological mapping of cardiac structures.

DateAug 18, 2011 (initiated Apr 29, 2011)
StatusTerminated
ReasonThe recall was expanded because between February and April 2011, Biosense Webster, Inc. notified customers that a voluntary Customer Notification had been Initiated due to the inaccessibility of the electronic version of the Instructions for Use (e-IFU) for the "10 ft. Interface Cable, 10 Pin - Shielded Tip Pins to 12 Pin - Blue", Catalog Number D128709, through the J&J Gateway (i.e., e-IFU websi
Root causeLabeling Change Control
ActionBiosense Webster sent an "URGENT FIELD SAFETY NOTICE" letter dated April 29, 2011 to all affected customers. The letter describes the product, problem, and the action taken by the firm. The letter notifies and advises customers that the updated IFUs for ISMUS Catheter with Auto ID Technology and the Duo-Decapolar catheters have been released and posted on the e-IFU website as of April 22, 2011 and April 21, 2011. Customers were instructed to complete the Acknowledgement Form and return the form to Biosense. Customers with questions or require additionial information were instructed to co
Quantity325 units
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K953768CORDIS WEBSTER A20 DIAGNOSTIC DEFLECTABLE TIP CATHETER
Cordis Webster, Inc., Nov 22, 1995