Recall Z-3268-2011 ยท LFR Glucose Dehydrogenase, Glucose

Bayer Healthcare, LLC

Contour, Contour TS Blood Glucose Test Strips and Bonus Packs Intended for self-testing by people with diabetes and healthcare professionals to monitor glucose concentrations in whole blood

DateSep 19, 2011 (initiated Aug 12, 2011)
StatusTerminated
ReasonAll 10- or 25- count glucose test strip vials, either loose vials or found in patient starter kits of Contour, Contour TS and Contour" USB blood glucose meters are being recalled a packaging issue which, under certain conditions, can cause some test strips packaged in some small-count vials to operate outside of US FDA-cleared performance specifications and produce a low-bias reading during blood
Root causePackaging
ActionBayer sent an "URGENT DIABETES TEST STRIP RECALL" letter dated August 12, 2011, to its customers with the name of the test strips, sizes recalled, reason for the recall, potential injury, and instructions to discontinue use and return the test strips to Bayer. The recall letter also included a recall response form that was to be faxed back to Bayer at 1-800-876-2243. For questions call Customer Logistics at 1-800-348-2637.
Quantity3,617,287 vials/ 54,555,460 test strips
DistributionNationwide Distribution and Puerto Rico
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K062058ASCENSIA CONTOUR BLOOD GLUCOSE METER, MODEL 7151; REAGENT STRIP, MODEL 7080
Bayer Healthcare, Sep 11, 2006