Recall Z-0187-2012 ยท LLI Bed, Therapeutic, Ac-Powered, Adjustable Home-Use
Invacare Corporation
Invacare 600 lb. Bariatric Bed, Head motor extension cable used on component numbers BAR5490 and BAR5490IVC. These components are part of the following Bed Model Number(s): BAR600, BARPKGCA, BAR600IVC, BARPKGIVC-1633, and BARPKG.
| Date | Nov 14, 2011 (initiated Jun 23, 2011) |
| Status | Terminated |
| Reason | Invacare Corporation decided to recall the products because risk analysis reveals that miskeying or disconnection of the head motor extension cable during bed set-up may result in product damage, or in rare instances, the potential for fire. |
| Root cause | Labeling design |
| Action | On 9/19/11, Invacare sent Recall Notificaiton Letters to their customers. They instructed their customers to quarantine the product, complete and return a response form. |
| Quantity | 35,747 |
| Distribution | Distribution Nationwide and to United Arab Emirates, Dominican Republic, Egypt, Guam, Mexico, & Virgin Islands. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.