Recall Z-2964-2011 ยท HJB Instrument, Measuring, Corneal Radius
Haag-Streit USA Inc
Lenstar LS 900 EyeSuite Biometry version i2.000, Four (4) Protective Cardboard Containers Per Each System. Haag-Streit USA, Inc. A non-invasive, non-contact OLCR (Optical Low Coherence Reflectometer) Biometer used for obtaining ocular measurements and performing calculations to assist in determination of the appropriate power and type of IOL (Intraocular Lens) for implantation after removal of th
| Date | Aug 9, 2011 (initiated Jun 20, 2011) |
| Status | Terminated |
| Reason | A patient record may be linked to an incorrect/inconsistent set of calculation results under specific circumstances when using EyeSuite Software version i2.000. |
| Root cause | Software change control |
| Action | Haag-Streit USA, Inc. sent an "Urgent Field Safety Notices" dated June 20, 2011 to all affected customers. The letter included a description of the affected product and problem. The firm instructed their customers to make sure all new and current templates are saved and upgrade to EyeSuite version i4.000. Customers are asked to complete and return the Software Update Verification Form. For additional information contact the Haag-Streit at (513) 398-3937. |
| Quantity | 216 |
| Distribution | Nationwide (USA) Distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K082891 | LENSTAR, MODEL LS900 Haag-Streit AG, Oct 20, 2009 |