Recall Z-2998-2011 ยท JAQ System, Applicator, Radionuclide, Remote-Controlled
Varian Medical Systems, Inc. Oncology Systems
Titanium Fletcher-style Applicator Set - Defined Geometry - CT Compatible with Titanium intrauterine probe with stopper length 30 mm, 30 degree angle. Varian Medical Systems, Palo Alto, CA; Manufactured by Varian Medical Systems Haan GmbH, Haan, Germany. To treat cancer of the uterus, cervix, endometrium and vagina.
| Date | Aug 11, 2011 (initiated Jul 13, 2011) |
| Status | Terminated |
| Reason | Potential of weld failure of the Titanium intrauterine probe at the cervical stopper. |
| Root cause | Device Design |
| Action | Varian Medical Systems sent an "URGENT MEDICAL DEVICE CORRECTION/URGENT FIELD SAFETY NOTICE" letter dated July 13, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The letter instructs customers to return all affected product to Varian Brachy Therapy by using the contact information provided in the letter. Replacement product will be provided free of charge to the customers. Customers should contact their local Varian Customer Support District or Regional Manager for questions regarding this notice. |
| Quantity | 437 - all sizes |
| Distribution | Worldwide Distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K983436 | GAMMAMED PLUS HIGH DOSE RATE REMOTE AFTERLOADING SYSTEM Gammamed USA, Oct 29, 1998 |