Recall Z-3055-2011 ยท JAQ System, Applicator, Radionuclide, Remote-Controlled

Varian Medical Systems, Inc. Oncology Systems

Flexible Probe with Blocking Washer (part of Segmented Cylinder Set) The Segmented Cylinder Applicator Set was developed to treat cancer of the vagina, vaginal stump and rectum.

DateAug 22, 2011 (initiated Jul 13, 2011)
StatusTerminated
ReasonBlocking washers on the Flexible Probe with Blocking washers may shift and cause potential for treatment length error.
Root causeDevice Design
ActionVarian Medical Systems sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated July 13, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The letter instructs customers to locate old design applicator probes and guiding tubes and to cease use of the affected products. Information in the letter included instructions on how to return the products to Varian BrachyTherapy. Customers should contact their local Varian Customer Support District or Regional Manager for questions regarding this notice.
Quantity1019 total
DistributionWorldwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K033371VARIAN INTRACAVITARY BRACHYTHERAPY APPLICATORS
Varian Medical Systems, Dec 24, 2003