ClearanceBrief
Recall Z-2991-2011 ยท JAQ System, Applicator, Radionuclide, Remote-Controlled

Varian Medical Systems, Inc. Oncology Systems

GammaMed Plastic Needle with Mandrin; 2.0mm diameter, length 113 mm. Remote controlled radionuclide applicator system. Varian Medical Systems, Inc. Mfg by Varian Medical Systems Haan GmbH, Haan, Germany Product Usage: The Interstitial Plastic Needles with 2 mm diameter are designed for interstitial treatments in areas such as head and neck, gynecology, breast or prostate. In order to avoid bend

DateAug 10, 2011 (initiated Jul 13, 2011)
StatusTerminated
ReasonNeedle tip may separate from the needle shaft such that the tip may detach in the patient upon withdrawal.
Root causeDevice Design
ActionVarian sent an "Urgent Medical Device Correction/Urgent Field Safety Notice" letter dated July 13, 2011 to all possibly affected customers. This notification supplements and modifies the previous "Customer Technical Bulletins" originally sent in September 2008. The initial Customer Technical Bulletins informed customers of the possibility that needle tips cans crack and separate. It also reminded users that the product has a limited life and to check needles prior to use as per instructions. Current correction notice letter instruct customers to stop using any plastic needles in inventory wi
Quantity601 - all sizes.
DistributionWorldwide Distribution - USA (nationwide) and the countries of: Argentina, Austria, Bangladesh, Belgium, Brazil, Canada, Columbia, Cyprus, Czech Republic Denmark, Dominican Republic, Estonia, France, Germany, Greece, Hungary, India, Japan, Kazahastan, Kuwait, Lebanon, Macedonia, Mexico, Moldova,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K983436GAMMAMED PLUS HIGH DOSE RATE REMOTE AFTERLOADING SYSTEM
Gammamed USA, Oct 29, 1998